Dpjo

PDF

In vitro flexural strength evaluation of a mini-implant prototype designed for Herbst appliance anchorage

Klaus BARRETTO-LOPES, Gladys Cristina DOMINGUEZ, André TORTAMANO, Jesualdo Luiz ROSSI, Julio Wilson VIGORITO

Aim: The purpose of this study was to evaluate the limit of flexural strength of a miniimplant prototype designed for Herbst appliance anchorage. Methods: After sample size calculation, four specimens with the new mini-implant were submitted to a single cantilever flexure test using a universal testing machine. The limit of flexural force strength was calculated. Results: The mini-implant prototype showed a limit of flexural force of 98.2 kgf, which was the lowest value found. Conclusion: The mini-implant prototype designed for Herbst appliance anchorage can withstand higher strength than the maximum human bite reported in the literature.

Keywords: Orthodontic appliances. Orthodontics. Herbst appliance. Mini-implant.

Friday, April 19, 2024 10:21